Determinants and clinical impact of pressure drift in manoscan anorectal high resolution manometry system
G Parthasarathy1, J McMaster1, K Feuerhak1
1Clinical Enteric Neuroscience Translational and Epidemiological Research (C.E.N.T.E.R.) Program, Division of Gastroenterology and Hepatology, Department of Medicine, Mayo Clinic, Rochester, MN, USA.
Background:
Pressure drift (PD), resulting from differences between room and body temperature, reduces the accuracy of pressure measurements with the Manoscan high resolution manometry (HRM) system. Our aims were to assess PD during anorectal HRM.
Methods:
Defined as the residual pressure measured immediately after the catheter was removed, PD was calculated for each sensor and averaged across all 12 sensors in 454 anorectal consecutive studies recorded with 3 HRM catheters. The relationship between PD and study duration, number of prior uses of a catheter, and peak and average pressure exposure during a study were evaluated. The correction of PD with a software algorithm (thermal compensation) was evaluated in 76 studies where the most distal sensor was outside the body.
Key Results:
The PD varied among sensors and across catheters. The average PD (7.3 ± 0.2 mmHg) was significantly greater for newer catheters, during longer studies, or when sensors were exposed to higher pressures. Together, these factors explained 81% of the variance in overall PD. After thermal compensation, the uncorrected median PD for the most distal sensor was 2.5-5 mmHg over the study duration. Correcting this changed the interpretation (e.g., as abnormal instead of normal) of at least 1 anorectal parameter in eight of 76 studies.
Conclusions & Inferences:
During anorectal HRM, PD declines with catheter use and is greater for newer catheters, when sensors are exposed to higher pressures, and for studies of longer duration. While PD is partially corrected with thermal compensation algorithms, the impact on interpretation is modest.
More Related Videos
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
04:54Invasive Hemodynamic Assessment for the Right Ventricular System and Hypoxia-Induced Pulmonary Arterial Hypertension in Mice
Published on: October 24, 2019
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement