Related Experiment Video
Updated: Mar 22, 2026

Evaluation of the Efficacy of Organic Peroxyacids for Eradicating Dairy Biofilms Using an Approach Combining Static and Dynamic Methods
Published on: December 9, 2022
In-Use Evaluation of Peracetic Acid for High-Level Disinfection of Endoscopes
Wu Chenjiao1, Zhang Hongyan, Gu Qing
1Wu Chenjiao, MD, is Nurse-in-Charge, Digestive Endoscopy Center, The First Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China. Zhang Hongyan, MD is Nurse-in-Charge, The First Teaching Hospital of Xinjiang Medical University Endoscopy Center, Wulumoqi, China. Gu Ging, MD is Associate Chief Nurse, Digestive Endoscopy Center, The First Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China. Zhong Xiaoqi, MD is Nurse-in-Charge, Digestive Endoscopy Center, The First Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China. Gu Liying, MD is Associate Chief Nurse, Digestive Endoscopy Center, The First Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China. Fang Ying, MD is Associate Chief Nurse, Digestive Endoscopy Center, The First Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China.
Abstract:
Many high-level disinfectants have been used for disinfection of endoscopes such as 2% glutaraldehyde (GA), 0.55% ortho-phthalaldehyde (OPA), and peracetic acid (PAA). Both GA and OPA are widely used in disinfection of endoscopes and have been previously discussed, but there is little research on the practical use of PAA as an endoscope disinfectant. An experimental model of a flexible gastrointestinal endoscope being contaminated with 9 strains of microorganism was designed. After the cleaning and disinfecting procedure was completed, we evaluated the biocidal activity (850 ppm PAA, 2% GA, and 0.55% OPA) on our flexible gastrointestinal endoscope model. We also evaluated sterilization effectiveness of PAA on other bacteria, including some antibiotic-resistant bacteria (methicillin-resistant Staphylococcus aureus, vancomycin-resistant Enterococcus, and Clostridium difficile). The residual bacterial colony count number of the PAA-disinfected endoscope was significantly lower than that of the GA- and OPA-disinfected endoscopes. The biocidal effect and efficiency of the endoscope disinfection by PAA appeared to be better than either the GA- or OPA-disinfected endoscope. PAA has demonstrated a good sterilization effect on other bacterial species; of particular note are common antibiotic-resistant bacteria such as methicillin-resistant Staphylococcus aureus, vancomycin-resistant Enterococcus, and Clostridium difficile. The results of this study demonstrate that PAA is a fast and effective high-level disinfectant for use in the reprocessing of flexible endoscopes.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Endoscopic Procedures V: ERCP
Patient...
Chemical Agents for Microbial Control
Endoscopic Procedures III: Video Capsule Endoscopy