Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Comparing SUV values of images at PET-CT console and the RT planning console using identical dataset of a study
Anusheel Munshi1, Sayan Paul, Biplab Sarkar
1Department of Radiation Oncology, Fortis Memorial Research Institute, Gurgaon, Haryana, India.
Purpose:
The use of positron emission tomography (PET) for radiotherapy planning purposes has become increasingly important in the last few years.In the current study, we compared the SUV values of images at the PET CT console to the SUV values obtained at the RT planning workstation.
Materials And Methods:
The PET-CT cylindrical body phantom was filled with a uniform 18F solution of 5.3. ± 0.27 kBq/mL radioactivity concentration. PET-CT scans were performed on a16 slice Time of Flight system. On a single day, the three consecutive scans were done at three time points 15 minutes apart to generate time points image data sets titled T1, T2, and T3. SUV calculations were performed by drawing region of interest. (ROI) encompassing the entire hot spot on each slice on the PET-CT console and the iPlan workstation. Minimum SUV, Maximum SUV and the Mean SUV were recorded. Statistical analysis was done using the SPSS software. (SPSS Inc.) (Version 18).
Results:
The absolute difference in average max SUV values i.e. Max (PET-CT) - Max (iPlan) for the time points T1, T2 and T3 were -0.168 (SD 0.175), -0.172 (SD 0.172) and -0.178 (SD 0.169). The difference in the minimum SUV values were -0.513 (SD 0.428), -0.311 (SD 0.358) and -0.303 (SD 0.322), respectively. Finally, the difference in the mean SUV values were -0.107 (SD 0.040), -0.096 (SD 0.067) and -0.072 (SD 0.044), respectively.
Conclusions:
Our study found out that the average difference in the two systems for maximum SUV values was < 0.2 absolute units.Our study suggests good reproducibility of SUV between the two systems. The relevance of these findings would be of seminal importance in current and future SUV-based PET-CT-based contouring in treatment planning systems.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET