Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose
Related Experiment Video
Updated: Mar 22, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Hae-Il Park1,2, Seong-Su Lee3, Jang-Won Son3
1Department of Laboratory Medicine, Bucheon St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Bucheon-si Gyeonggi-do, Republic of Korea.
Background:
Element™ Auto-coding Blood Glucose Monitoring System (BGMS; Infopia Co., Ltd., Anyang-si, Korea) was developed for self-monitoring of blood glucose (SMBG).
Methods:
Precision, linearity, and interference were tested. Eighty-four capillary blood samples measured by Element™ BGMS were compared with central laboratory method (CLM) results in venous serum. Accuracy was evaluated using ISO 15197:2013 criteria.
Results:
Coefficients of variation (CVs; mean) were 2.4% (44.2 mg/dl), 3.7% (100.6 mg/dl), and 2.1% (259.8 mg/dl). Linearity was shown at concentrations 39.25-456.25 mg/l (y = 0.989 + 0.984x, SE = 17.63). Up to 15 mg/dl of galactose, ascorbic acid, and acetaminophen, interference > 10.4% was not observed. Element™ BGMS glucose was higher than CLM levels by 3.2 mg/dl (at 200 mg/dl) to 8.2 mg/dl (at 100 mg/dl). The minimum specification for bias (3.3%) was met at 140 and 200 mg/l glucose. In the Clarke and consensus error grids, 100% of specimens were within zone A and B. For Element™ BGMS values, 92.9% (78/84) to 94.0% (79/84) were within a 15 mg/dl (< 100 mg/dl) or 15% (> 100 mg/dl) of the average CLM value.
Conclusion:
Element™ BGMS was considered an appropriate SMBG for home use; however, the positive bias at low-to-mid glucose levels requires further improvement.
Related Concept Videos
Automated Microbial Diagnostics
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...