Effects of gravity-induced upper-limb blood pressure changes on wave transmission and arterial radial waveform
Giacomo Pucci1, Francesca Battista, Fabio Anastasio
1aDipartimento di Medicina, Università degli Studi di Perugia, Perugia bUnità di Medicina Interna, Azienda Ospedaliero-Universitaria di Terni, Terni, Italy cYazmonit Ltd., Eshtaol, Israel dFaculty of Medicine and Health Sciences, Macquarie University, Sydney, Australia.
Background:
Local blood pressure (BP) changes induced by arm tilting may influence pressure wave transmission and reflection. We investigated the effects of upper-limb tilting on radial augmentation index (rAIx) and related central measures [aortic augmentation index (aAIx)].
Methods:
In 45 volunteers (age 49 ± 19 years), supine brachial BP and radial artery waveforms were obtained by applanation tonometry with the dominant arm stretched and gently supported in three different positions: at the heart level, with the BP cuff 15 cm above heart level (approximately +30°), and 15 cm below heart level (-30°).
Results:
Brachial SBP/DBP was 120/68 ± 17/8 mmHg. Mean arterial pressure changed predictably with arm tilting (99 ± 12 mmHg at -30°, 88 ± 10 mmHg at 0°, 77 ± 11 mmHg at +30°, all P < 0.001). rAIx decreased at -30° (69 ± 22%), and increased at +30° (93 ± 20%) compared with 0° (82 ± 20%, all P less than 0.001). Changes in rAIx (value at +30° minus value at -30°) showed an inverse relationship with age (r = -0.32, P = 0.03). Heart rate, BP and rAIx did not change in the contralateral arm, which was held at the heart level during the examination. aAIx followed the same pattern as rAIx (123 ± 27% at -30°, 144 ± 33% at +30°, 136 ± 31% at 0°, all P less than 0.001); changes in rAIx and aAIx were strongly related each other (r = 0.82, P < 0.001).
Conclusion:
Acute gravitational upper-limb BP changes generate opposite, profound changes in rAIx, and major artifactual changes in aAIx. These findings provide a rationale for recommending to keep the upper limb at the heart level during radial waveform assessment.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Blood Pressure