Related Experiment Video
Updated: Mar 22, 2026

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Investigating the reproducibility of a complex multifocal radiosurgery treatment
M Niebanck1, T Juang1, J Newton1
1Duke University Medical Center, Durham, NC, USA.
Abstract:
Stereotactic radiosurgery has become a widely used technique to treat solid tumors and secondary metastases of the brain. Multiple targets can be simultaneously treated with a single isocenter in order to reduce the set-up time to improve patient comfort and workflow. In this study, a 5-arc multifocal RapidArc treatment was delivered to multiple PRESAGE® dosimeters in order to explore the repeatability of the treatment. The three delivery measurements agreed well with each other, with less than 3% standard deviation of dose in the target. The deliveries also agreed well with the treatment plan, with gamma passing rates greater than 90% (5% dose-difference, and 2 mm distance-to-agreement criteria). The optical-CT PRESAGE® system provided a reproducible measurement for treatment verification, provided measurements were made immediately following treatment.