Related Experiment Video
Updated: Mar 22, 2026

Complete and Partial Aortic Occlusion for the Treatment of Hemorrhagic Shock in Swine
Published on: August 24, 2018
A side-hole catheter for catheterization of a difficult internal mammary artery
Jung Suk Oh1, Byung Gil Choi1, Ho Jong Chun1
1Department of Radiology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea.
Insights
A new side-hole catheter offers a 100% success rate for internal mammary artery (IMA) catheterization in patients with hemoptysis, overcoming challenges posed by anatomical variations. This technique improves safety and efficiency in embolization procedures.
Area of Science:
- Interventional Radiology
- Vascular Surgery
- Pulmonary Medicine
Background:
- The internal mammary artery (IMA) is a potential source of hemoptysis in chronic lung disease patients.
- Catheterizing the IMA presents challenges due to anatomical variations, potentially increasing contrast and radiation exposure.
Purpose of the Study:
- To assess the safety and efficiency of a novel side-hole catheter for internal mammary artery (IMA) catheterization.
- To evaluate this new technique in patients experiencing hemoptysis.
Main Methods:
- A retrospective review of 96 transarterial embolization procedures involving the IMA.
- Comparison of a novel side-hole catheter against conventional methods in 17 cases (18%) where selective IMA angiography failed.
- Utilized a modified cobra-type curved catheter with a side hole for micro-wire and catheter passage.
Main Results:
- The novel side-hole catheter achieved a 100% technical success rate in difficult IMA catheterizations.
- Conventional methods failed in 18% of cases due to vascular tortuosity or abnormal IMA takeoff.
- Catheterization time was significantly reduced with the side-hole catheter (2 min) compared to the cobra catheter (17 min).
- No procedure-related complications were reported.
Conclusions:
- The side-hole catheter technique is effective for challenging internal mammary artery (IMA) access in hemoptysis patients.
- Further design improvements are recommended to enhance the speed and ease of IMA catheterization and angiography.
Abstract:
Background The internal mammary artery (IMA) can be a source of hemoptysis in patients with chronic lung disease. Intervention via the IMA can be a challenge due to anatomic variations and lead to excessive contrast use and radiation exposure. Purpose To evaluate safety and efficiency of a new side-hole catheter for the catheterization of the IMA in patients with hemoptysis. Material and Methods From January 2011 to August 2014, a total of 96 transarterial embolization procedures required exact evaluation of the IMA due to chronic lung disease involving the anterior thorax. In 17 cases (18%) of these 96 procedures, the conventional selective IMA angiography failed and instead a novel side-hole catheter as a modification of a cobra-type curved catheter was used. The side hole allowed passage of a micro-wire and catheter. Results Failed catheterizations were due to severe vascular tortuosity, acutely angulated subclavian artery, or abnormal takeoff of the IMA. The Cobra shaped catheter with the microcatheter through the side-hole catheter yielded a technical success rate of 100%. Longer time was required to catheterize the IMA with the Cobra shaped catheter than with the side-hole catheter (17 vs. 2 min, P < 0.05). There were no procedure-related complications. Conclusion Side-hole catheter technique is useful in patients whose internal mammary artery is difficult to access. Further design revisions are needed to improve the ease and speed of IMA catheterization and angiography.