A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
The "Pain wave": P‑wave oversensing in subcutaneous ICD
Badder Kattih1, David Duncker1, Thorben König1
1Department of Cardiology and Angiology, Hannover Medical School, Carl-Neuberg-Str.1, 30625, Hannover, Deutschland.
A patient with a subcutaneous implantable defibrillator experienced inappropriate shocks due to oversensing. Continuous monitoring of sensing vectors is crucial for managing implantable cardioverter defibrillator (ICD) performance.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Subcutaneous implantable defibrillators (S-ICDs) offer an alternative to transvenous devices for sudden cardiac arrest prevention.
- Idiopathic ventricular fibrillation necessitates device implantation for arrhythmia management.
Observation:
- A 46-year-old male patient developed inappropriate shocks from his S-ICD.
- The event was attributed to P- and T-wave oversensing, leading to triple counting.
Findings:
- Despite advanced discrimination algorithms like INSIGHT(TM), oversensing remains a challenge.
- The incidence of oversensing-related shocks can be reduced but not entirely eliminated.
Implications:
- Regular and thorough evaluation of sensing vectors during various patient activities is essential.
- Optimizing S-ICD sensing performance requires ongoing vigilance and patient monitoring.
- This case highlights the need for continuous assessment of S-ICD function post-implantation.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...