Importance of Calibration Method in Central Blood Pressure for Cardiac Structural Abnormalities
Kazuaki Negishi1, Hong Yang2, Ying Wang2
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Tasmania, Australia. Kazuaki.Negishi@utas.edu.au.
Background:
Central blood pressure (CBP) independently predicts cardiovascular risk, but calibration methods may affect accuracy of central systolic blood pressure (CSBP). Standard central systolic blood pressure (Stan-CSBP) from peripheral waveforms is usually derived with calibration using brachial SBP and diastolic BP (DBP). However, calibration using oscillometric mean arterial pressure (MAP) and DBP (MAP-CSBP) is purported to provide more accurate representation of true invasive CSBP. This study sought to determine which derived CSBP could more accurately discriminate cardiac structural abnormalities.
Methods:
A total of 349 community-based patients with risk factors (71±5years, 161 males) had CSBP measured by brachial oscillometry (Mobil-O-Graph, IEM GmbH, Stolberg, Germany) using 2 calibration methods: MAP-CSBP and Stan-CSBP. Left ventricular hypertrophy (LVH) and left atrial dilatation (LAD) were measured based on standard guidelines.
Results:
MAP-CSBP was higher than Stan-CSBP (149±20 vs. 128±15mm Hg, P < 0.0001). Although they were modestly correlated (rho = 0.74, P < 0.001), the Bland-Altman plot demonstrated a large bias (21mm Hg) and limits of agreement (24mm Hg). In receiver operating characteristic (ROC) curve analyses, MAP-CSBP significantly better discriminated LVH compared with Stan-CSBP (area under the curve (AUC) 0.66 vs. 0.59, P = 0.0063) and brachial SBP (0.62, P = 0.027). Continuous net reclassification improvement (NRI) (P < 0.001) and integrated discrimination improvement (IDI) (P < 0.001) corroborated superior discrimination of LVH by MAP-CSBP. Similarly, MAP-CSBP better distinguished LAD than Stan-CSBP (AUC 0.63 vs. 0.56, P = 0.005) and conventional brachial SBP (0.58, P = 0.006), whereas Stan-CSBP provided no better discrimination than conventional brachial BP (P = 0.09).
Conclusions:
CSBP is calibration dependent and when oscillometric MAP and DBP are used, the derived CSBP is a better discriminator for cardiac structural abnormalities.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement