Effectiveness of the implementation of a simple radiation reduction protocol in the catheterization laboratory
Alfonso Jurado-Román1, Ignacio Sánchez-Pérez1, Fernando Lozano Ruíz-Poveda1
1Unidad de Hemodinámica, Servicio de Cardiología, Hospital General Universitario de Ciudad Real.
Background And Purpose:
A reduction in radiation doses at the catheterization laboratory, maintaining the quality of procedures is essential. Our objective was to analyze the results of a simple radiation reduction protocol at a high-volume interventional cardiology unit.
Methods:
We analyzed 1160 consecutive procedures: 580 performed before the implementation of the protocol and 580 after it. The protocol consisted in: the reduction of the number of ventriculographies and aortographies, the optimization of the collimation and the geometry of the X ray tube-patient-receptor, the use of low dose-rate fluoroscopy and the reduction of the number of cine sequences using the software "last fluoroscopy hold".
Results:
There were no significant differences in clinical baseline features or in the procedural characteristics with the exception of a higher percentage of radial approach (30.7% vs 69.6%; p<0.001) and of percutaneous coronary interventions of chronic total occlusions after the implementation of the protocol (2.1% vs 6.7%; p=0,001). Angiographic success was similar during both periods (98.3% vs 99.2%; p=0.2). There were no significant differences between both periods regarding the overall duration of the procedures (26.9 vs 29.6min; p=0.14), or the fluoroscopy time (13.3 vs 13.2min; p=0.8). We observed a reduction in the percentage of procedures with ventriculography (80.9% vs 7.1%; p<0.0001) or aortography (15.4% vs 4.4%; p<0.0001), the cine runs (21.8 vs 6.9; p<0.0001) and the dose-area product (165 vs 71 Gyxcm(2); p<0.0001).
Conclusions:
With the implementation of a simple radiation reduction protocol, a 57% reduction of dose-area product was observed without a reduction in the quality or the complexity of procedures.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization IV: Nursing Management
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization III: Left Heart Catheterization