Manpreet S Mundi1, Lisa Epp1, Ryan T Hurt1,2,3,4
11 Division of Endocrinology, Diabetes, Metabolism and Nutrition, Mayo Clinic, Rochester, Minnesota, USA.
Background:
Enteral tube misconnections resulting in enteral nutrition being provided into the wrong port or tube has been increasingly reported, prompting the development of new international design standards for medical device tubing connectors.
Methods:
Four sample enteral feeds with varying viscosity were placed into a syringe with either the current feeding connector or a prototype ENFit connector. The force (N) required to compress the syringe was gathered through a dynamic mechanical analyzer (RSA G2 Solids Analyzer; TA Instruments).
Results:
The force needed to compress the syringe was lowest with a fiber-containing formula (Jevity 1 Cal, Abbott Nutrition, Columbus, OH), and a decline in force was detected with the ENFit compared with the current connector (8.61 ± 0.27 N vs 9.62 ± 0.23 N, P < .001). A commercial blenderized formula (Liquid Hope, Functional Formularies, Centerville, OH) required slightly higher force with the ENFit than the current connector (16.82 ± 1.23 N vs 14.5 ± 0.03 N, P < .001). Another commercial blenderized formula (Salmon, Oats, & Squash, Real Food Blends, Chesterton, IN) and a homemade blenderized recipe required significantly more force with the ENFit than the current tube (34.12 ± 0.95 N vs 22.91 ± 0.06 N, P < .001; 34.95 ± 0.06 N vs 27.72 ± 0.07 N, P < .001, respectively).
Conclusion:
ENFit adapter implementation is the largest change to home enteral nutrition in more than a decade. Additional rigorous testing is required to ensure that the needs of all patients receiving home enteral nutrition are met prior to widespread adoption.
More Related Videos
08:32Adapting Human Videofluoroscopic Swallow Study Methods to Detect and Characterize Dysphagia in Murine Disease Models
Published on: March 1, 2015
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
Related Concept Videos
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Enteral Nutrition II: Nasointestinal and Gastrostomy Feeding
Nasointestinal Feeding
Nasointestinal feeding involves placing a tube...
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Drug Delivery: Enteral Route
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...