Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Mid-Infrared Pumped Laser-Induced Thermal Grating Spectroscopy for Detection of Acetylene in the Visible Spectral
Anna-Lena Sahlberg1, Johannes Kiefer2, Marcus Aldén1
1Division of Combustion Physics, Lund University, Sweden.
Abstract:
We present mid-infrared laser-induced thermal grating spectroscopy (IR-LITGS) using excitation radiation around 3 µm generated by a simple broadband optical parametric oscillator (OPO). Acetylene as a typical small hydrocarbon molecule is used as an example target species. A mid-infrared broadband OPO pumped by the fundamental output of a neodymium-doped yttrium aluminum garnet (Nd:YAG) laser was used to generate the pump beams, with pulse energies of 6-10 mJ depending on the wavelength. The line width of the OPO idler beam was ∼5 cm(-1), which is large enough to cover up to six adjacent acetylene lines. The probe beam was the radiation of a 532 nm cw solid state laser with 190 mW output power. Signals were generated in atmospheric pressure gas flows of N2, air, CO2 and Ar with small admixtures of C2H2 A detection limit of less than 300 ppm was found for a point measurement of C2H2 diluted in N2 As expected, the oscillation frequency of the IR-LITGS signal was found to have a large dependency on the buffer gas, which allows determination of the speed of sound. Moreover, the results reveal a very strong collisional energy exchange between C2H2 and CO2 compared to the other gases. This manifests as significant local heating. In summary, the MIR-LITGS technique enables spectroscopy of fundamental vibrational transitions in the infrared via detection in the visible spectral range.
More Related Videos
Related Concept Videos
Flame Photometry: Overview
Gas Chromatography: Types of Detectors-II
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
IR and UV–Vis Spectroscopy of Carboxylic Acids
However, the stretching absorptions for the C=O bond vary depending on the structure of carboxylic acids. The C=O bond of the free carboxylic acids shows a higher stretching frequency, 1760 cm−1, while H-bonded carboxylic acids (dimers) exhibit stretching absorptions at a lower frequency,...
Atomic Emission Spectroscopy: Lab