Background:
Nearly 14% to 39% TKA patients report dissatisfaction causing incomplete return of function. We proposed that the kinematics of knees implanted with patient-specific prostheses using patient-specific cutting guides would be closer to normal.
Methods:
Eighteen matched cadaver lower limbs were randomly assigned to two groups: group A was implanted with patient-specific implants using patient-specific cutting guides; group B, the contralateral knee, was implanted with a standard design using intramedullary alignment cutting guides. Knee kinematics were measured on a dynamic closed-kinetic-chain Oxford knee rig, simulating a deep knee bend and in a passive rig testing varus-valgus laxity.
Results:
The difference from normal kinematics was lower for group A compared to group B for active femoral rollback, active tibiofemoral adduction, and for passive varus-valgus laxity.
Conclusions:
Our results support the hypothesis that knees with patient-specific implants generate kinematics more closely resembling normal knee kinematics than standard knee designs.
Clinical Relevance:
Restoring normal kinematics may improve function and patient satisfaction after total knee arthroplasty.