Related Experiment Videos
Quantitative analysis of trazodone hydrochloride in tablets by an ion-selective electrode
H Suzuki1, K Akimoto, H Nakagawa
1Pharmaceuticals Research Center, Kanebo, Ltd., Osaka, Japan.
Abstract:
A simple assay procedure for the quantitation of trazodone hydrochloride in tablets, without prior separation, has been developed using a trazodone-selective electrode. The electrode was based on a trazodone:tetrakis(p-chlorophenyl)borate ion-pair complex, dioctyl phthalate, and polyvinyl chloride matrix that were mounted on a PTFE membrane. The electrode showed a near-Nernstian response in the range of 10(-2) to 5 x 10(-6) M trazodone over the pH range of 2.5 to 5.0, with a cationic slope of 58 mV/concentration decade. The durability of the electrode and the reproducibility of the performance among the electrodes were sufficient. The electrode was used for the determination of the pKa value of trazodone. The selectivity of the electrode to a number of interferences was investigated. Many inorganic cations (alkali and alkaline earth metals) and pharmaceutical excipients did not interfere, but some organic ammonium compounds interfered according to their extractability. Determination of 10 to 3000 micrograms/mL of trazodone hydrochloride in aqueous solution showed an average recovery of 100.6% (mean standard deviation 0.4%) by direct potentiometry. This assay was applied to determine trazodone hydrochloride in tablets and the results compared favorably with those obtained by high-performance liquid chromatography.