Comparison Between Doppler-Echocardiography and Uncalibrated Pulse Contour Method for Cardiac Output Measurement: A

Sabino Scolletta1, Federico Franchi, Stefano Romagnoli

  • 11Department of Medical Biotechnologies, Anesthesiology and Intensive Care, University Hospital of Siena, Siena, Italy. 2Department of Anesthesiology and Critical Care, Azienda Ospedaliero-Universitaria Careggi, University of Florence, Florence, Italy. 3Department of Health Science, University of Florence, Florence, Italy. 4Department of Emergency and Critical Medicine, Anesthesiology and Intensive Care, Mugello Hospital, Florence, Italy. 5Dipartimento di Scienze Biomediche e Sanità Pubblica, Università Politecnica delle Marche, AOU Ospedali Riuniti di Ancona, Clinica di Anestesia e Rianimazione, Torrette di Ancona, Italy. 6Dipartimento di Emergenza-Urgenza, Azienda Sanitaria Massa Carrara, Ospedale di Pontremoli, Pontremoli, Italy. 7Department of Surgery, Intensive Care Unit IV, Azienda Ospedaliera Universitaria Pisana, Pisa, Italy. 8Department of Anesthesia and Surgical Critical Care, Hospital Universitari i Politécnic La Fe, Valencia, Spain. 9Department of Emergency and Critical Medicine, Anesthesiology and Intensive Care, Santa Maria Annunziata Hospital, Florence, Italy. 10Department of Emergency and Critical Medicine, Anesthesiology and Intensive Care, San Giovanni di Dio Hospital, Florence, Italy. 11Department of Cardiothoracic Anesthesia and Intensive Care, San Raffaele Hospital, Istituto Scientifico San Raffaele, Milan, Italy. 12Department of Anaesthesia, Heart and Lung Centre, Royal Wolverhampton NHS Trust, Wolverhampton, United Kingdom. 13Department of Anesthesiology and Intensive Care, University of Foggia, Foggia, Italy. 14Department of Cardiovascular, Respiratory, Nephrologic, Anesthesiologic and Geriatric Sciences, University of Rome "La Sapienza," Rome, Italy. 15Emergency Department, Istituto di Anestesia e Rianimazione, Azienda Ospedaliera Universitaria di Padova, Padova, Italy. 16Department of Anesthesiology and Critical Care, Lariboisière Hospital, Assistance Publique-Hôpitaux de Paris and UMR INSERM 1160: Alloimmunité, Autoimmunité, Transpla

Critical Care Medicine
|April 21, 2016
PubMed

Insights

The pulse contour method MostCare effectively estimates cardiac output in critically ill patients, offering a viable alternative to echocardiography. This study compared MostCare with echocardiography in 400 ICU patients, finding a significant correlation and acceptable error margins.

Area of Science:

  • Critical Care Medicine
  • Cardiovascular Physiology
  • Medical Technology Assessment

Background:

  • Cardiac output (CO) monitoring is crucial in intensive care units (ICUs).
  • Echocardiography provides noninvasive CO estimation, while pulse contour analysis offers a less invasive approach.
  • Comparing different CO monitoring methods is essential for clinical practice optimization.

Purpose of the Study:

  • To compare the accuracy of the pulse contour method (MostCare) with Doppler echocardiography for cardiac output estimation.
  • To evaluate MostCare's performance in a large, unselected population of critically ill patients across various clinical conditions.

Main Methods:

  • A prospective, multicenter observational study involving 400 critically ill patients in 15 European ICUs.
  • Cardiac output was measured using both Doppler echocardiography and the MostCare system.
  • Statistical analyses included Bland-Altman and linear regression to assess agreement and correlation.

Main Results:

  • A significant correlation (r = 0.85, p < 0.0001) was observed between MostCare and echocardiography CO measurements.
  • The overall mean bias was -0.03 L/min with 95% limits of agreement of -1.54 to 1.47 L/min.
  • The overall percentage error was 27.3%, with lower errors noted in sepsis (24%) and trauma (26%) patient categories.

Conclusions:

  • The MostCare system demonstrates potential as a reliable alternative to echocardiography for cardiac output assessment in diverse ICU patient populations.
  • The findings support the integration of pulse contour analysis for continuous and less invasive hemodynamic monitoring in critical care settings.
Abstract

Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.9K
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.8K
Pulse01:05

Pulse

The pulse is one of the most fundamental physiological indicators of the body's cardiovascular health. It is the rhythmic expansion and contraction of the arterial walls in response to the pressure generated by the heart's pumping action.
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
4.6K
Pulse01:16

Pulse

When the heart pumps blood out, arterial elastic fibers play a crucial role in sustaining a high-pressure gradient. They expand to accommodate the received blood and then recoil - a process known as the pulse that can be either manually palpated or electronically quantified. Despite a reduction in its effect with increased distance from the heart, elements of the pulse's systolic and diastolic components persist, observable even at the arteriole level.
The pulse serves as a clinical...
2.6K
Cardiac Output I:Effect of Heart Rate on Cardiac Output01:19

Cardiac Output I:Effect of Heart Rate on Cardiac Output

Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
3.2K
Imaging Studies for Cardiovascular System I:Echocardiography01:17

Imaging Studies for Cardiovascular System I:Echocardiography

Cardiac imaging studies encompass a wide range of noninvasive and minimally invasive techniques designed to visualize the heart's structure and function in detail. One such technique is echocardiography, which uses high-frequency ultrasound waves to produce detailed images of the heart, known as echocardiograms.
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
896