Related Experiment Video
Updated: Mar 22, 2026

Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Validation of device algorithm to differentiate pacemaker-mediated tachycardia from tachycardia due to atrial
1Department of Medicine, Division of Cardiology, Cornell University Medical Center, New York, New York.
Background:
Current cardiac devices cannot always differentiate between pacemaker-mediated tachycardia (PMT) and tracking of sinus or atrial tachycardia. We previously derived a novel algorithm for distinguishing the 2 mechanisms based on the specific termination response to postventricular atrial refractory period extension, atrial rates, and changes in atrial electrogram morphology.
Objective:
The purpose of this study was to evaluate how this algorithm would have performed in a clinical setting based on previously recorded PMT events.
Methods:
We applied our algorithm to a database of 122 de-identified stored electrograms that were classified as PMT by 43 remotely monitored devices.
Results:
Of the 122 events stored as "PMT," 3 episodes were excluded because the device recording was consistent with atrial fibrillation. Of the remaining 119 episodes, our algorithm was able to correctly reclassify 92 events (77%) as tracking of sinus or atrial tachycardia rather than true PMT. The VAV response following postventricular atrial refractory period extension, which is specific to tracking of atrial or sinus tachycardia, was seen in 72% of these cases. Changes in atrial rate and atrial electrogram morphology were able to reclassify the remainder of episodes. Finally, we observed that 12 of 83 episodes (14%) misclassified as PMT in cardiac resynchronization devices resulted in loss of cardiac biventricular pacing.
Conclusion:
Applying a novel diagnostic algorithm to current cardiac devices improves the proper diagnosis of true PMT rather than tracking of atrial or sinus tachycardia. Enhanced accuracy of diagnosis reduces the likelihood of inappropriate clinical decisions.
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Mechanism of Cardiac Arrhythmias
Dysrhythmias V: Evaluating Dysrhythmias