An Evaluation of the Pea Pod System for Assessing Body Composition of Moderately Premature Infants
Elisabet Forsum1, Elisabeth Olhager2, Caroline Törnqvist3
1Department of Clinical and Experimental Medicine, Linköping University, SE-581 83 Linköping, Sweden. elisabet.forsum@liu.se.
Insights
The Pea Pod device accurately assesses body composition in groups of moderately premature infants. Further research is recommended to validate its use in diverse premature infant populations.
Area of Science:
- Neonatalogy
- Pediatric Nutrition
- Body Composition Analysis
Background:
- Assessing growth quality in premature infants is crucial for optimal nutrition.
- The Pea Pod device, utilizing air displacement plethysmography, measures infant body composition.
- Limited evaluation of the Pea Pod method exists for premature infants.
Purpose of the Study:
- To evaluate the accuracy of the Pea Pod device for assessing body composition in premature infants.
- To compare Pea Pod estimates with a three-component model reference in this population.
Main Methods:
- Body composition was assessed in 14 premature infants (3-7 days old, 32-35 weeks gestation).
- Measurements included body weight, volume, fat-free mass density (Pea Pod), and total body water (isotope dilution).
- A three-component model provided reference estimates for fat-free mass density and body composition.
Main Results:
- Pea Pod-predicted fat-free mass density showed no significant difference from reference estimates (p > 0.05).
- Body fat percentage assessed by Pea Pod was not significantly different from reference values.
- Biological variability in fat-free mass density was minimal (0.55% of the average value).
Conclusions:
- The Pea Pod system demonstrates accuracy for assessing body composition in groups of newborn, moderately premature infants.
- Additional studies are necessary to further validate the Pea Pod system's utility in broader premature infant populations.
- Suggestions for future research directions in this area are provided.
Abstract:
(1) BACKGROUND: Assessing the quality of growth in premature infants is important in order to be able to provide them with optimal nutrition. The Pea Pod device, based on air displacement plethysmography, is able to assess body composition of infants. However, this method has not been sufficiently evaluated in premature infants; (2) METHODS: In 14 infants in an age range of 3-7 days, born after 32-35 completed weeks of gestation, body weight, body volume, fat-free mass density (predicted by the Pea Pod software), and total body water (isotope dilution) were assessed. Reference estimates of fat-free mass density and body composition were obtained using a three-component model; (3) RESULTS: Fat-free mass density values, predicted using Pea Pod, were biased but not significantly (p > 0.05) different from reference estimates. Body fat (%), assessed using Pea Pod, was not significantly different from reference estimates. The biological variability of fat-free mass density was 0.55% of the average value (1.0627 g/mL); (4) CONCLUSION: The results indicate that the Pea Pod system is accurate for groups of newborn, moderately premature infants. However, more studies where this system is used for premature infants are needed, and we provide suggestions regarding how to develop this area.