The Accuracy of Pulse Spectroscopy for Detecting Hypoxemia and Coexisting Methemoglobin or Carboxyhemoglobin
Axel Kulcke1, John Feiner, Ingolf Menn
1From the *Senspec, Rostock, Germany; †Department of Anesthesia and Perioperative Care, University of California San Francisco, San Francisco, California; and ‡Sentec, Therwil, Switzerland.
Background:
Pulse spectroscopy is a new noninvasive technology involving hundreds of wavelengths of visible and infrared light, enabling the simultaneous quantitation of multiple types of normal and dysfunctional hemoglobin. We evaluated the accuracy of a first-generation pulse spectroscopy system (V-Spec™ Monitoring System, Senspec, Germany) in measuring oxygen saturation (SpO2) and detecting carboxyhemoglobin (COHb) or methemoglobin (MetHb), alone or simultaneously, with hypoxemia.
Methods:
Nineteen volunteers were fitted with V-Spec probes on the forehead and fingers. A radial arterial catheter was placed for blood sampling during (1) hypoxemia with arterial oxygen saturations (SaO2) of 100% to 58.5%; (2) normoxia with MetHb and COHb increased to approximately 10%; (3) 10% COHb or MetHb combined with hypoxemia with SaO2 of 100% to 80%. Standard measures of pulse-oximetry performance were calculated: bias (pulse spectroscopy measured value - arterial measured value) mean ± SD and root-mean-square error (Arms).
Results:
The SpO2 bias for SaO2 approximately 60% to 100% was 0.06% ± 1.30% and Arms of 1.30%. COHb bias was 0.45 ± 1.63, with an Arms of 1.69% overall, and did not degrade substantially during moderate hypoxemia. MetHb bias was 0.36 ± 0.80 overall and stayed small with hypoxemia. Arms was 0.88 and was <3% at all levels of SaO2 and MetHb. Hypoxemia was also accurately detected by pulse spectroscopy at elevated levels of COHb. At elevated MetHb levels, a substantial negative bias developed, -10.3 at MetHb >10%.
Conclusions:
Pulse spectroscopy accurately detects hypoxemia, MetHb, and COHb. The technology also accurately detects these dysfunctional hemoglobins during hypoxemia. Future releases of this device may have an improved SpO2 algorithm that is more robust with methemoglobinemia.
More Related Videos
07:07Tumor Hypoxia Assessment: In Vivo 3D Oxygen Imaging Through Electron Paramagnetic Resonance
Published on: February 14, 2025
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Oxygen Transport in the Blood
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...