Setup and validation of a convenient sampling procedure to promptly and effectively stabilize vitamin C in blood and
Barbara Rossi1, Francesca Tittone1, Simonetta Palleschi2
1Department of Hematology, Oncology and Molecular Medicine, Istituto Superiore di Sanità, Viale Regina Elena, 299, 00161, Rome, Italy.
Abstract:
Vitamin C (ascorbic acid, AA) is very labile in nature and decays quickly after blood withdrawal. To ensure AA stability, the current procedure prescribes immediate plasma acidification followed by sample storage at ultra-low temperature. The aim of this study was to set up a pre-analytical procedure to promptly stabilize AA at routine temperatures while minimizing both specimen manipulation and instrumental requirement. Blood from healthy subjects was collected in lithium-heparin gel separator tubes containing or not different reducing agents (dithioerythritol, tris(2-carboxyethyl)phosphine, n-acetylcysteine and sodium thiosulfate). Plasma AA stability during blood and plasma storage at routine temperatures was evaluated. Plasma AA concentration was assayed by RP-HPLC-UV under ion suppression conditions. Each of the reductants tested was able to slow down the ex vivo degradation of plasma AA; dithioerythritol was the most effective. Five to 10 mmol/L dithioerythritol did not interfere with blood separation and allowed plasma AA to be stabilized up to 6 h, 24 h and 60 days at room temperature, +4 °C and -25 °C, respectively. The method worked well even in case of delayed blood separation and/or incomplete vacutainer filling. The procedure is feasible and reliable. Of special usefulness in clinical and epidemiological studies, prompt plasma manipulation after blood withdrawal or special storage equipments are not required. Graphical Abstract Collecting blood in tubes containing a reducing agent is a feasible method to promptly and effectively stabilize plasma vitamin C at routine temperature.
Related Concept Videos
Convenience Sampling Method
Convenience sampling is a non-random method of sample selection; this method selects individuals that are easily accessible and may result in biased data. For example, a marketing...
Effects of Temperature on Free Energy
Composition of Blood Plasma
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Protein Buffers in Blood Plasma and Cells
Certain amino acids can exist in a zwitterion state at a...
Vitamins