Tracheal Stenosis and Cuff Pressure: Comparison of Minimal Occlusive Volume and Palpation Techniques
Ziae Totonchi1, Fatemeh Jalili1, Seyed Mohammadreza Hashemian2
1Rajaei Cardiovascular Medical & Research Center, Iran University of Medical Sciences, Tehran, Iran.
Background:
Tracheal stenosis, which has received more emphasis recently, is a common post intubation complication and may develop due to different reasons. One important reason is the endotracheal tube cuff pressure. Therefore, this study sought to examine the accuracy of diagnostic test for palpation and minimal occlusive volume techniques to measure the endotracheal tube cuff pressure.
Materials And Methods:
In this cross sectional study, the accuracy of diagnostic tests for palpation and minimal occlusive volume techniques to measure the endotracheal tube cuff pressure was assessed in 101 patients aged over 18 years who had undergone open heart surgery and post-surgical mechanical ventilation in the ICU.
Results:
In the palpation technique, the cuff pressure of 27 patients (26.7%) was reported to be out of the permissible range and for the rest of them (74 patients, 73.3%) it was within the permissible range. Then, the cuff pressure was checked by the standard method using a manometer and after comparing the results it was found that the cuff pressure of 92 patients (91.1%) was not in the permissible range and only nine patients (8.9%) had a cuff pressure within the permissible range (20-30 cm H2O). In minimal occlusive volume method compared with the standard method, 22 patients (21.7%) had cuff pressure within the permissible range of 20-30 cm H2O, and 79 of them (78.2%) had cuff pressure out of the permissible range and higher than the upper limit.
Conclusion:
This study recommends that the best way to measure the endotracheal tube cuff pressure is to use a cuff manometer, and when it is not available, the minimal occlusive volume would be a better alternative compared to the palpation technique to keep the cuff pressure within a proper range to avoid tracheotomy complications such as tracheal stenosis.
Related Concept Videos
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...