Related Experiment Video
Updated: Mar 22, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Efficiency Calibration for Environmental Gamma Spectrometry Using GATE
1*Deptartment of Physics, Faculty of Science, Kuwait University, Khaldia, Kuwait.
Abstract:
This work investigated the utility of performing efficiency calibration for environmental gamma spectrometry using the Monte Carlo based, free of charge GATE toolbox. The validity of this approach was tested by comparing output efficiency values of an in-house developed GATE-based program with experimental measurements covering various geometries and primary photon energies. The results of this comparison revealed relative deviations within ±20%, thus validating the employed computational approach. Moreover, the GATE-based method was able to predict quantities that are generally difficult to measure experimentally, such as the number of interactions preceding full energy absorption. These computationally obtained predictions were found to be in agreement with theory.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...