Determination of (226)Ra in produced water by liquid scintillation counting
José Marcus Godoy1, Lucas M Vianna1, Maria Luiza D P Godoy2
1Department of Chemistry, Pontifícia Universidade Católica do Rio de Janeiro, Rua Marquês de São Vicente, 225 - Gávea, 22451-900 Rio de Janeiro, Brazil.
Abstract:
It is proposed a method for the determination of (226)Ra in offshore platform liquid effluent samples (produced water). The method is based on a two-phase liquid scintillation counting system and allows for the direct and simple determination of (226)Ra content. Samples with high barium content may also have high (226)Ra concentration. Therefore, the sample volume is based on the barium concentration and ranges from 10 mL to 100 mL. Our new method was tested using multiple real samples and was compared with the BaSO4 precipitation method. The results based on the LSC were 30% higher than the precipitation method, which is attributed to the self-absorption of alpha particles in the BaSO4 precipitate. The determination of both (226)Ra and (228)Ra in the liquid effluent of offshore oil platforms is mandatory in Brazil. Thus, a second method of accurately assessing (228)Ra content remains necessary.