Related Experiment Video
Updated: Mar 22, 2026

Novel Triple-Loop Technique for Suturing TFCC Injuries without Transosseous Tunnel
Published on: May 23, 2025
Foreign body reaction after trapeziectomy and Dacron interposition
Philippe Willekens1, Koenraad Verstraete2, Nadine Hollevoet1
1Department of Orthopaedic Surgery and Traumatology, Ghent University Hospital, De Pintelaan 185, 9000 Ghent, Belgium.
Dacron interposition arthroplasty for trapeziometacarpal osteoarthritis can cause severe foreign body reactions, including bone erosion. Due to these risks, this implant is no longer used, and patients require careful monitoring.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Rheumatology
Background:
- Trapeziometacarpal osteoarthritis is a common condition affecting the thumb base.
- Dacron interposition arthroplasty has been used as a treatment option.
- Foreign body reactions to implants can lead to complications.
Purpose of the Study:
- To evaluate patients treated with trapeziectomy and Dacron interposition.
- To describe the incidence and characteristics of foreign body reactions associated with Dacron implants.
Main Methods:
- Retrospective review of 10 patients (11 hands) who underwent trapeziectomy and Dacron interposition arthroplasty between 2004 and 2010.
- Clinical examination, DASH score, VAS for pain and satisfaction, grip, and pinch strength assessment.
- Radiographic and histological analysis for foreign body reactions.
Main Results:
- Two patients (2 hands) required implant removal due to foreign body reactions, confirmed histologically with giant cells.
- Seven out of nine remaining hands showed radiological signs of foreign body reaction, including bone erosion.
- Three patients experienced severe reactions; mean DASH score was 32, VAS for pain 1.6, and satisfaction 8.8.
Conclusions:
- Dacron interposition arthroplasty is associated with a significant incidence of foreign body reactions, including bone erosion.
- The use of Dacron implants in this context has been discontinued.
- Close monitoring of patients with existing Dacron implants is recommended.
Related Concept Videos
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History: