A Comparison of Three C-Arm Draping Techniques to Minimize Contamination of the Surgical Field
Grigory E Gershkovich1, Nathan C Tiedeken, David Hampton
1Department of Orthopaedic Surgery, Einstein Medical Center, Philadelphia, PA.
Abstract:
The use of intraoperative fluoroscopy has become a routine and useful adjunct within orthopaedic surgery. However, the fluoroscopy machine may become an additional source of contamination in the operating room, particularly when maneuvering from the anterior-posterior position to the lateral position. Consequently, draping techniques were developed to maintain sterility of the operative field and surgeon. Despite a variety of methods, no studies exist to compare the sterility of these techniques specifically when the fluoroscopy machine is in the lateral imaging position. We evaluated the sterility of 3 c-arm draping techniques in a simulated operative environment. The 3 techniques consisted of a traditional 3-quarter sterile sheet attached to the side of the operative table, a modified clip-drape method, and a commercially available sterile pouch. Our study demonstrated that the traditional method poses a high risk for sterile field contamination, whereas the modified clip-drape method and commercially available sterile pouch kept floor contamination furthest from the surgical field. With the current data, we urge surgeons to use modified techniques rather than the traditional draping method.
Insights
Sterile draping techniques for intraoperative fluoroscopy (c-arm) in orthopaedic surgery are crucial. Modified methods, unlike traditional ones, significantly reduce sterile field contamination during lateral imaging.
Area of Science:
- Orthopaedic Surgery
- Infection Control
- Surgical Technology
Background:
- Intraoperative fluoroscopy (c-arm) is standard in orthopaedic surgery.
- C-arm manipulation, especially to lateral positions, risks operating room contamination.
- Existing draping techniques lack comparative sterility data for lateral c-arm use.
Purpose of the Study:
- To evaluate and compare the sterility of three c-arm draping techniques.
- To identify optimal methods for maintaining sterility during lateral fluoroscopy in simulated surgical environments.
Main Methods:
- Simulation of an operative environment to test draping techniques.
- Evaluation of three distinct c-arm draping methods: traditional 3-quarter sheet, modified clip-drape, and a commercial sterile pouch.
- Assessment of sterile field integrity and floor contamination levels.
Main Results:
- The traditional 3-quarter sterile sheet method showed a high risk of sterile field contamination.
- Both the modified clip-drape method and the commercial sterile pouch effectively minimized floor contamination.
- These modified techniques kept contamination furthest from the surgical field compared to the traditional method.
Conclusions:
- Modified c-arm draping techniques are superior to the traditional method for maintaining sterility.
- Surgeons should adopt modified clip-drape or commercial sterile pouch methods.
- Improved draping practices are essential for reducing infection risk in orthopaedic procedures using fluoroscopy.
More Related Videos
08:26Clinical Application of Single-Surgeon, Three-Port, Laparoscopic Resection for Colorectal Cancer with Natural Orifice Specimen Extraction
Published on: March 24, 2023
08:34Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
Related Concept Videos
Hand hygiene
Hand washing...
Asepsis
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
PPE Use in Healthcare Settings II: Doffing
Handwashing III: During the Procedure and Post-Procedure Steps
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...