Related Experiment Video
Updated: Mar 21, 2026

Assessment of Kidney Function in Mouse Models of Glomerular Disease
Published on: June 30, 2018
Development and validation of an enzyme-linked immunosorbent assay for the quantification of gelonin in mouse plasma
Frank A Engler1, Joseph P Balthasar1
1a Department of Pharmaceutical Sciences , School of Pharmacy and Pharmaceutical Sciences, University at Buffalo - The State University of New York , Buffalo , New York , USA.
Abstract:
This article details the development and validation of an enzyme-linked immunosorbent assay (ELISA) for the quantification of gelonin in mouse plasma. The ELISA was validated for intra- and inter-day variability and for accuracy over a standard curve range of 7.5-100 ng/mL. The assay was then applied to assess gelonin pharmacokinetics in mice. Results from the ELISA were compared to data obtained from a parallel study conducted with (125)Iodine-labeled gelonin, with quantification via gamma counting. The ELISA demonstrated good precision, as the percent coefficient of variation of quality control samples in intra-day and inter-day validation ranged from 5.4-9.3% and 2.9-7.3%, respectively. Sample recoveries ranged from 98.3-105% of nominal values. The ELISA method yielded lower plasma concentrations of gelonin than found from the less-specific gamma counting method. Consequently, pharmacokinetic analyses yielded significantly higher estimates for volume of distribution (106 ± 31 vs. 55.8 ± 13 mL/kg) and plasma clearance (34.7 ± 6.6 vs. 10.9 ± 2.1 mL/min/kg) for data determined by ELISA vs. by gamma counting.