Polarized hard X-ray photoemission system with micro-positioning technique for probing ground-state symmetry of strongly correlated materials
Related Experiment Video
Updated: Mar 21, 2026

Experimental Methods for Spin- and Angle-Resolved Photoemission Spectroscopy Combined with Polarization-Variable Laser
Published on: June 28, 2018
Hidenori Fujiwara1, Sho Naimen1, Atsushi Higashiya2
1Division of Materials Physics, Graduate School of Engineering Science, Osaka University, Toyonaka, Osaka 560-8531, Japan.
Abstract:
An angle-resolved linearly polarized hard X-ray photoemission spectroscopy (HAXPES) system has been developed to study the ground-state symmetry of strongly correlated materials. The linear polarization of the incoming X-ray beam is switched by a transmission-type phase retarder composed of two diamond (100) crystals. The best value of the degree of linear polarization was found to be -0.96, containing a vertical polarization component of 98%. A newly developed low-temperature two-axis manipulator enables easy polar and azimuthal rotations to select the detection direction of photoelectrons. The lowest temperature achieved was 9 K, offering the chance to access the ground state even for strongly correlated electron systems in cubic symmetry. A co-axial sample monitoring system with long-working-distance microscope enables the same region on the sample surface to be measured before and after rotation. Combining this sample monitoring system with a micro-focused X-ray beam by means of an ellipsoidal Kirkpatrick-Baez mirror (25 µm × 25 µm FWHM), polarized valence-band HAXPES has been performed on NiO for voltage application as resistive random access memory to demonstrate the micro-positioning technique and polarization switching.