Related Experiment Video
Updated: Mar 21, 2026

Monitoring the Wall Mechanics During Stent Deployment in a Vessel
Published on: May 8, 2012
Evaluation of cover effects on bare stent mechanical response
D J McGrath1, B O'Brien1, M Bruzzi1
1Biomechanics Research Centre (BMEC), Biomedical Engineering, College of Engineering and Informatics, NUI Galway, Ireland.
Abstract:
Covered tracheobronchial stents are used to prevent tumour growth from reoccluding the airways. In the present work a combination of experimental and computational methods are used to present the mechanical effects that adhered covers can have on stent performance. A prototype tracheobronchial stent is characterised in bare and covered configurations using radial force, flat plate and a novel non-uniform radial force test, while computational modelling is performed in parallel to extensively inform the physical testing. Results of the study show that cover configuration can have a significant structural effect on stent performance, and that stent response (bare or covered) is especially loading specific, highlighting that the loading configuration that a stent is about to be subjected to should be considered before stent implantation.