Related Experiment Video
Updated: Mar 21, 2026

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Internal Jugular Vein Avulsion Complicating Dual-Lumen VV-ECMO Cannulation: An Unreported Complication of Avalon
Christopher L Tarola1, A Dave Nagpal1
1Division of Cardiac Surgery, Department of Surgery, University Hospital, London Health Sciences Center, London, Ontario, Canada.
Abstract:
The Avalon Elite dual lumen cannula (Avalon Laboratories, Rancho Dominguez, CA) was developed as a means of single-site cannulation for veno-venous extracorporeal membrane oxygenation in patients with severe refractory respiratory failure. This cannulation strategy has several advantages over conventional strategies, however, is not without complication. We describe a case of right internal jugular vein "on a stick" (avulsion) after cannulation with the Avalon Elite cannula, and corroborate the critical importance of imaging guidance when using this system.
Related Concept Videos
Hemodialysis II: Procedure and Complications
Venous Thrombosis III: Interprofessional Care
Hemodialysis I: Introduction
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies