Influence of monte carlo variance with fluence smoothing in VMAT treatment planning with Monaco TPS
B Sarkar1, A Manikandan, M Nandy
1Department of Radiation Oncology, AMRI Hospitals, Kolkata, India.
Introduction:
The study aimed to investigate the interplay between Monte Carlo Variance (MCV) and fluence smoothing factor (FSF) in volumetric modulated arc therapy treatment planning by using a sample set of complex treatment planning cases and a X-ray Voxel Monte Carlo-based treatment planning system equipped with tools to tune fluence smoothness as well as MCV.
Materials And Methods:
The dosimetric (dose to tumor volume, and organ at risk) and physical characteristic (treatment time, number of segments, and so on) of a set 45 treatment plans for all combinations of 1%, 3%, 5% MCV and 1, 3, 5 FSF were evaluated for five carcinoma esophagus cases under the study.
Result:
Increase in FSF reduce the treatment time. Variation of MCV and FSF gives a highest planning target volume (PTV), heart and lung dose variation of 3.6%, 12.8% and 4.3%, respectively. The heart dose variation was highest among all organs at risk. Highest variation of spinal cord dose was 0.6 Gy.
Conclusion:
Variation of MCV and FSF influences the organ at risk (OAR) doses significantly but not PTV coverage and dose homogeneity. Variation in FSF causes difference in dosimetric and physical parameters for the treatment plans but variation of MCV does not. MCV 3% or less do not improve the plan quality significantly (physical and clinical) compared with MCV greater than 3%. The use of MCV between 3% and 5% gives similar results as 1% with lesser calculation time. Minimally detected differences in plan quality suggest that the optimum FSF can be set between 3 and 5.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:20MRI-guided dmPFC-rTMS as a Treatment for Treatment-resistant Major Depressive Disorder
Published on: August 11, 2015
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...