Related Experiment Video
Updated: Mar 21, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Technical Note: Dosimetry of Leipzig and Valencia applicators without the plastic cap
D Granero1, C Candela-Juan2, J Vijande3
1Department of Radiation Physics, ERESA, Hospital General Universitario, Valencia 46014, Spain.
Forgetting the plastic cap during high dose rate (HDR) brachytherapy significantly increases skin dose. Always verify cap placement before HDR skin treatments to prevent dangerous overdosage.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- High dose rate (HDR) brachytherapy is common for skin lesions.
- Applicators often use a plastic cap to ensure electronic equilibrium and shield skin from secondary electrons.
Purpose of the Study:
- To investigate the dosimetric consequences of omitting the plastic cap during HDR brachytherapy fraction delivery.
- This impact has not been previously studied in the literature.
Main Methods:
- geant4 Monte Carlo simulations were used to model Leipzig and Valencia applicators.
- Simulations were performed with and without the plastic cap.
- Radiochromic film measurements validated the simulation results.
Main Results:
- Absence of the cap caused a 1500% dose increase within 1 mm of the skin for Leipzig applicators.
- A 180% dose increase was observed for Valencia applicators under the same conditions.
- Deeper than 1 mm, the dose increase diminished to approximately 10%.
Conclusions:
- Omitting the plastic cap leads to significant skin overdosage.
- Verification of cap placement before HDR skin treatments is crucial.
- Inclusion of cap verification in timeout checklists is recommended.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Radiation: Applications
The average...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET