Related Experiment Videos
High-performance liquid chromatography of sialooligosaccharides and gangliosides
Z el Rassi1, C Horváth, R K Yu
1Department of Chemical Engineering, Yale University, New Haven, CT 06520.
Abstract:
Glycans were cleaved from gangliosides and separated by high-performance liquid chromatography (HPLC). The columns were packed with bonded stationary phases made of microparticulate, macroporous silica with serotonin, phenylpropanolamine or tryptamine as the biogenic amine ligate. The ganglioside oligosaccharides were eluted in the order of increasing number of sialic acid residues in the molecule and their retention decreased with the ionic strength of the mobile phase. Best selectivity was obtained in the pH range from 3.0 to 4.0. The two major sialic acids, N-acetylneuraminic and N-glycolylneuraminic acids, were separated by lectin affinity chromatography using an HPLC column packed with silica-bound wheat germ agglutinin and 10 mM phosphate buffer, pH 4.0, as the eluent. Throughout this study, isocratic elution was used and the column effluent was monitored at 195 nm.