Related Experiment Video
Updated: Mar 21, 2026

Implantation and Control of Wireless, Battery-free Systems for Peripheral Nerve Interfacing
Published on: October 20, 2021
Inhibition and Excitation of Bladder Function by Tibial Nerve Stimulation Using a Wirelessly Powered Implant: An
Zainab Moazzam1, Austin R Duke2, Paul B Yoo3
1Institute of Biomaterials and Biomedical Engineering, University of Toronto, Toronto, Ontario, Canada.
Purpose:
Tibial nerve stimulation is a minimally invasive neuromodulation treatment of overactive bladder. However, in addition to our limited understanding of the underlying mechanisms, there are also questions regarding the long-term delivery of tibial nerve stimulation therapy in patients. We aimed to characterize the effects of stimulation frequency using a wirelessly powered implantable stimulation device.
Methods And Materials:
Six α-chloralose anesthetized adult male cats were used in this study. A multicontact lead was surgically implanted subcutaneously in the hind limb and used to stimulate the tibial nerve. Using an isovolumetric bladder a short duration of electrical pulses was applied at amplitudes 3 times the motor threshold and at frequencies from 2 to 20 Hz.
Results:
Implant driven stimulation of the tibial nerve resulted in frequency dependent activation of bladder reflexes. Low frequency tibial nerve stimulation (2 Hz) consistently evoked excitatory responses (mean ± SE 32.9% ± 3.8%). In contrast, higher frequency tibial nerve stimulation (6 to 20 Hz) inhibited bladder function (overall mean 14.9% ± 2.4%). Although low foot motor thresholds were achieved at initial implantation (mean 0.83 ± 0.05 mA), a notable elevation in threshold amplitude was observed 5 hours after implantation.
Conclusions:
To our knowledge this study provides the first evidence of frequency dependent modulation of bladder function in anesthetized cats. The inhibitory influence of tibial nerve stimulation at frequencies above 6 Hz transitioned to an excitatory effect at 2 Hz. Taken together these preclinical data support the feasibility of using a wirelessly powered implantable device to potentially modulate bladder function in patients.