Optimizing contrast media application in coronary CT angiography at lower tube voltage: Evaluation in a circulation
Madeleine Kok1, Casper Mihl1, Babs M F Hendriks2
1Department of Radiology, Maastricht University Medical Center, The Netherlands; CARIM, School for Cardiovascular Diseases, Maastricht University Medical Center, Maastricht, The Netherlands.
Purpose:
The purpose was to investigate optimal contrast media (CM) injection parameters for lower kVp settings, whilst maintaining diagnostic attenuation levels.
Methods And Materials:
First, a circulation phantom with physiological parameters (BP 120/80mmHg, HR 60bpm) was used. A fixed CM injection protocol was used for each kVp setting (300mgI/ml [Iopromide], volume=45ml, flow rate=6.0ml/s, iodine delivery rate [IDR]=1.8gI/s, iodine load=13.5gI; at 120, 100, 80 and 70kVp). Then, IDR was decreased by steps of 0.2gI/s for each kVp setting, until diagnostically insufficient attenuation values were reached (<325HU). In order to keep injection time constant (7.5s), total iodine load (TIL) was reduced accordingly. Second, clinical applicability at 120 and 100kVp was evaluated in patients (n=60) referred for coronary CT angiography. A standard and reduced (12% less) CM protocol was used based on weight classes and scan duration ('high-pitch': 1s; 'adaptive sequence' and 'helical': 7s). Attenuation levels of the coronary arteries were measured and compared between protocols.
Results:
Using a fixed CM injection at each kVp level resulted in the following HU values: 335HU±31 (120kVp); 425HU±30 (100kVp); 587HU±29 (80kVp); 666HU±27 (70kVp). Keeping diagnostic enhancement levels (353HU±28) CM could be reduced as follows: 12% for 100kVp; 45% for 80kVp and 56% for 70kVp. Diagnostic enhancement levels could be reproduced with concurrent CM reduction (-12% at 100kV) in the clinical setting (382HU±35).
Conclusion:
CM injection parameters can be substantially reduced at low kVp settings (up to 56% at 70kVp), whilst maintaining diagnostic attenuation levels. This may play an important role in CT imaging of the coronary arteries as well as cerebral and peripheral circulations in the future.
More Related Videos
10:16Retrograde Perfusion and Filling of Mouse Coronary Vasculature as Preparation for Micro Computed Tomography Imaging
Published on: February 10, 2012
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System IV: CMRI