Stent Lumen Visibility in Single-energy CT Angiography: Does Tube Potential Matter?
Johannes Boos1, Patric Kröpil1, Rotem Shlomo Lanzman1
1Department of Diagnostic and Interventional Radiology, Medical Faculty, University Dusseldorf, Moorenstraße 5, 40225 Dusseldorf, Germany.
Rationale And Objectives:
There has been a trend toward lowering tube potential in computed tomography angiography (CTA) examinations to reduce radiation dose or contrast medium dose. The aim of this study was to evaluate the influence of tube potential on peripheral artery in-stent lumen visibility in CTA examinations.
Materials And Methods:
Nine different peripheral artery stents were placed in a vessel phantom (inner diameter: 5 mm, surrounded by water) and scanned consecutively using a 128-row CT scanner with 70, 80, 100, 120, and 140 kV and two different concentrations of contrast medium to simulate contrast-enhanced blood. Medium-smooth and ultra-sharp reconstruction kernels with filtered back projection (B30f, B46f) and iterative reconstruction technique (I30f, I46f) were used. Visible in-stent lumen diameter and artifact width were evaluated using a semiautomatic software tool. All stents were scanned with digital angiography, which was regarded as the reference standard.
Results:
Averaged over all stents, visible in-stent lumen diameter ranged from 1.30 ± 0.21 mm (CM2/70 kV/I30f) to 3.13 ± 0.32 mm (CM1/120 kV/I46f). In-stent lumen diameters were significantly higher for 120 and 140 kV compared to 70 kV (2.39 ± 0.73 and 2.39 ± 0.66 mm vs 1.99 ± 0.69 mm; P = 0.01 and P = 0.005). Ultra-sharp reconstruction kernels lead to significantly better in-stent lumen visibility than smooth reconstruction kernels (B46f: 2.74 ± 0.34 mm vs B30f: 1.57 ± 0.36 mm; P < 0.001, respectively). Furthermore, in-stent lumen visibility was improved for iterative reconstructions compared to filtered back projection (I46f: 2.93 ± 0.30 mm vs B46f: 2.74 ± 0.34 mm; P < 0.001). Contrast medium concentration did not influence in-stent lumen visibility.
Conclusions:
Despite all known benefits of low kV CTA protocols, the use of a very low tube potential may hamper in-stent lumen visibility. A sharp kernel may be of value when evaluating the inner lumen of vascular stents.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT