Related Experiment Video
Updated: Mar 21, 2026

An Ultrasonic Tool for Nerve Conduction Block in Diabetic Rat Models
Published on: October 20, 2017
Modeling, Production, and Testing of an Echogenic Needle for Ultrasound-Guided Nerve Blocks
Paul E Bigeleisen1, Aaron Hess2, Richard Zhu3
1Department of Anesthesiology, University of Maryland School of Medicine, Baltimore, Maryland USA pbigeleisen@anes.umm.edu.
Abstract:
We have designed, produced, and tested an echogenic needle based on a sawtooth pattern where the height of the tooth was 1.25 times the wavelength of the ultrasound transducer. A numeric solution to the time-independent wave equation (Helmholtz equation) was used to create a model of backscattering from a needle. A 21-gauge stainless steel prototype was manufactured and tested in a water bath. Backscattering from the needle was compared to theoretical predications from our model. Based on these results, an 18-gauge prototype needle was fabricated from stainless steel and tested in a pig cadaver. This needle was compared to a commercial 18-gauge echogenic needle (Pajunk Medical Systems, Tucker, GA) by measuring the brightness of the needle relative to the background of sonograms of a needle in a pig cadaver. The backscattering from the 21-gauge prototype needle reproduced the qualitative predictions of our model. At 30° and 45° of insonation, our prototype performed equivalently to the Pajunk needle. At 60°, our prototype was significantly brighter than the Pajunk needle (P = .017). In conclusion, we chose a model for the design of an echogenic needle and modeled it on the basis of a solution to the Helmholtz equation. A prototype needle was tested in a water bath and compared to the model prediction. After verification of our model, we designed an 18-gauge needle, which performed better than an existing echogenic needle (Pajunk) at 60° of insonation. Our needle will require further testing in human trials.
More Related Videos
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Local Anesthetics: Differential Sensitivity of Nerve Fibers