Related Experiment Video
Updated: Mar 21, 2026

Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Research: Comparison of the Accuracy of a Pocket versus Standard Pulse Oximeter
Background:
Pulse oximetry has become an essential tool in clinical practice. With patient self-management becoming more prevalent, pulse oximetry self-monitoring has the potential to become common practice in the near future. This study sought to compare the accuracy of two pulse oximeters, a high-quality standard pulse oximeter and an inexpensive pocket pulse oximeter, and to compare both devices with arterial blood co-oximetry oxygen saturation.
Methods:
A total of 95 patients (35.8% women; mean [±SD] age 63.1 ± 13.9 years; mean arterial pressure was 92 ± 12.0 mmHg; mean axillar temperature 36.3 ± 0.4°C) presenting to our hospital for blood gas analysis was evaluated. The Bland-Altman technique was performed to calculate bias and precision, as well as agreement limits. Student's t test was performed.
Results:
Standard oximeter presented 1.84% bias and a precision error of 1.80%. Pocket oximeter presented a bias of 1.85% and a precision error of 2.21%. Agreement limits were -1.69% to 5.37% (standard oximeter) and -2.48% to 6.18% (pocket oximeter).
Conclusion:
Both oximeters presented bias, which was expected given previous research. The pocket oximeter was less precise but had agreement limits that were comparable with current evidence. Pocket oximeters can be powerful allies in clinical monitoring of patients based on a self-monitoring/efficacy strategy.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse
The pulse serves as a clinical...