Related Experiment Video
Updated: Mar 21, 2026

Neuronavigation and Laparoscopy Guided Ventriculoperitoneal Shunt Insertion for the Treatment of Hydrocephalus
Published on: October 14, 2022
Shunt Testing In Vivo: Observational Study of Problems with Ventricular Catheter
Zofia H Czosnyka1, Rohitiwa Sinha1, James A D Morgan1
1Division of Neurosurgery, Department of Clinical Neurosciences, University of Cambridge, Cambridge, UK.
None:
Most shunt obstructions happen at the inlet of the ventricular catheter. Three hundred six infusion studies from 2007 to 2011 were classified as having a typical pattern of either proximal occlusion or patency. We describe different patterns of shunt ventricular obstruction.Solid block: Cerebrospinal fluid (CSF) aspiration was impossible. Baseline pressure was without pulse waveform (respiratory waveform may be visible). A quick increase of pressure to a level compatible with the shunt's setting was recorded in response to infusion. Distal occlusion of the shunt via transcutaneous compression resulted in a rapid increase in pressure to levels above 50 mmHg. This pattern was attributed to a solid ventricular block.Slit ventricles: At baseline, a pattern similar to that of the solid block was observed. After compression, the pressure increases, the pulse waveform appears, and the intracranial pressure is often stabilized at 25-40 mmHg. It is probable that previously slit ventricles were opened during the test.Partial block: In a partial block of the ventricular catheter by an in-growing choroid plexus, the pulse waveform at baseline was observed and CSF aspiration was possible. During infusion, the pressure increased, but the pulse amplitude disappeared. During the increase in the pressure in the shunt prechamber, the connection with the ventricles is disturbed by repositioning of the plexus.Infusion study via the shunt prechamber is able to visualize ventricular obstruction of the hydrocephalus shunt.
More Related Videos
09:23Shunt Surgery, Right Heart Catheterization, and Vascular Morphometry in a Rat Model for Flow-induced Pulmonary Arterial Hypertension
Published on: February 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Shunt Admittances
To model this effect, the method of images is employed. This method involves replacing the Earth with an image conductor that mirrors the original...