Are simple IMRT beams more robust against MLC error? Exploring the impact of MLC errors on planar quality assurance
Jiazhou Wang1, Xiance Jin, Jiayuan Peng
1Fudan University Shanghai Cancer Center; Shanghai Medical College, Fudan University. wjiazhou@gmail.com.
Abstract:
This study investigated the impact of beam complexities on planar quality assur-ance and plan quality robustness by introducing MLC errors in intensity-modulate radiation therapy. Forty patients' planar quality assurance (QA) plans were enrolled in this study, including 20 dynamic MLC (DMLC) IMRT plans and 20 static MLC (SMLC) IMRT plans. The total beam numbers were 150 and 160 for DMLC and SMLC, respectively. Six different magnitudes of MLC errors were introduced to these beams. Gamma pass rates were calculated by comparing error-free fluence and error-induced fluence. The plan quality variation was acquired by comparing PTV coverage. Eight complexity scores were calculated based on the beam flu-ence and the MLC sequence. The complexity scores include fractal dimension, monitor unit, modulation index, fluence map complexity, weighted average of field area, weighted average of field perimeter, and small aperture ratio (< 5 cm2 and < 50cm2). The Spearman's rank correlation coefficient was calculated to analyze the correlation between these scores and gamma pass rate and plan quality varia-tion. For planar QA, the most significant complexity index was fractal dimension for DMLC (p = -0.40) and weighted segment area for SMLC (p = 0.27) at low magnitude MLC error. For plan quality, the most significant complexity index was weighted segment perimeter for DMLC (p = 0.56) and weighted segment area for SMLC (p= 0.497) at low magnitude MLC error. The sensitivity of planar QA was weakly associated with the field complexity with low magnitude MLC error, but the plan quality robustness was associated with beam complexity. Plans with simple beams were more robust to MLC error.
More Related Videos
09:45Large-area Scanning Probe Nanolithography Facilitated by Automated Alignment and Its Application to Substrate Fabrication for Cell Culture Studies
Published on: June 12, 2018
11:47Characterization of Surface Modifications by White Light Interferometry: Applications in Ion Sputtering, Laser Ablation, and Tribology Experiments
Published on: February 27, 2013
Related Concept Videos
Design of Prismatic Beams for Bending
Prismatic Beams: Problem Solving
The design begins with analyzing the beam as a free body to identify moments and force balances, thereby determining support reactions. Next, the...
Common Leveling Mistakes and Errors
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Deformation of a Beam under Transverse Loading
The insights from the bending moment diagram extend to...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...