Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
Published on: April 27, 2016
Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
Paula Rodriguez-Miguelez1, Nichole Seigler1, Ryan A Harris2
1Division of Clinical and Translational Sciences, Georgia Prevention Institute, Georgia Regents University.
Insights
The flow-mediated dilation (FMD) test assesses vascular endothelial function, crucial for understanding cardiovascular disease. This article standardizes the FMD protocol to improve accuracy and reproducibility in clinical settings.
Area of Science:
- Cardiology
- Vascular Biology
- Diagnostic Imaging
Background:
- Cardiovascular disease is a leading global cause of mortality and disability.
- Endothelial dysfunction, an imbalance in vasoactive substances, is implicated in atherosclerosis.
- Accurate assessment of endothelial function aids in understanding cardiovascular pathologies.
Purpose of the Study:
- To describe a standardized protocol for the flow-mediated dilation (FMD) test.
- To minimize confounding factors affecting FMD measurement precision and reproducibility.
Main Methods:
- The flow-mediated dilation (FMD) test involves forearm occlusion and reactive hyperemia.
- Ultrasound is used to assess brachial artery vasodilation.
- A standardized protocol is detailed to address physiological and technical issues.
Main Results:
- The FMD test is a widely used, non-invasive method for assessing endothelial function.
- FMD measurements are associated with future cardiovascular events.
- Standardization aims to enhance the precision and reproducibility of FMD assessments.
Conclusions:
- Standardizing the FMD protocol is essential for reliable endothelial function assessment.
- Improved FMD methodology can aid in understanding cardiovascular disease etiology.
- This standardized approach supports the clinical utility of FMD testing.
Abstract:
Cardiovascular disease is the primary cause of mortality and a major cause of disability worldwide. The dysfunction of the vascular endothelium is a pathological condition characterized mainly by a disruption in the balance between vasodilator and vasoconstrictor substances and is proposed to play an important role in the development of atherosclerotic cardiovascular disease. Therefore, a precise evaluation of endothelial function in humans represents an important tool that could help better understand the etiology of multiple cardio-centric pathologies. Over the past twenty-five years, many methodological approaches have been developed to provide an assessment of endothelial function in humans. Introduced in 1989, the FMD test incorporates a forearm occlusion and subsequent reactive hyperemia that promotes nitric oxide production and vasodilation of the brachial artery. The FMD test is now the most widely utilized, non-invasive, ultrasonic assessment of endothelial function in humans and has been associated with future cardiovascular events. Although the FMD test could have clinical utility, it is a physiological assessment that has inherited several confounding factors that need to be considered. This article describes a standardized protocol for determining FMD including the recommended methodology to help minimize the physiological and technical issues and improve the precision and reproducibility of the assessment.
More Related Videos
08:42Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
09:40Ultrasound Assessment of Flow-Mediated Dilation of the Brachial and Superficial Femoral Arteries in Rats
Published on: November 3, 2016
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):