Performance of cone-beam computed tomography and multidetector computed tomography in diagnostic imaging of the
Simon Veldhoen1,2, Maximilian Schöllchen3, H Hanken3
1Department of Diagnostic and Interventional Radiology, University Medical Center Hamburg, Martinistrasse 52, D-20246, Hamburg, Germany. veldhoen_s@ukw.de.
Purpose:
To compare multidetector computed tomography (MDCT) and cone-beam computed tomography (CBCT) regarding radiation, resolution, image noise, and image quality.
Methods:
CBCT and 256-MDCT were compared based on three scan protocols: Standard-dose (≈24 mGy), reduced-dose (≈9 mGy), and low-dose (≈4 mGy). MDCT images were acquired in standard- and high-resolution mode (HR-MDCT) and reconstructed using filtered back projection (FBP) and iterative reconstruction (IR). Spatial resolution in linepairs (lp) and objective image noise (OIN) were assessed using dedicated phantoms. Image quality was assessed in scans of 25 cadaver heads using a Likert scale.
Results:
OIN was markedly higher in FBP-MDCT when compared to CBCT. IR lowered the OIN to comparable values in standard-mode MDCT only. CBCT provided a resolution of 13 lp/cm at standard-dose and 11 lp/cm at reduced-dose vs. 11 lp/cm and 10 lp/cm in HR-MDCT. Resolution of 10 lp/cm was observed for both devices using low-dose settings. Quality scores of MDCT and CBCT did not differ at standard-dose (CBCT, 3.4; MDCT, 3.3-3.5; p > 0.05). Using reduced- and low-dose protocols, CBCT was superior (reduced-dose, 3.2 vs. 2.8; low dose, 3.0 vs. 2.3; p < 0.001).
Conclusion:
Using the low-dose protocol, the assessed CBCT provided better objective and subjective image quality and equality in resolution. Similar image quality, but better resolution using CBCT was observed at higher exposure settings.
Key Points:
• The assessed CBCT device provided better image quality at lower doses. • Objective and subjective image quality were comparable using higher exposure settings. • CBCT showed superior spatial resolution in standard-dose and reduced-dose settings. • Modern noise-reducing tools are used in CBCT devices currently. • MDCT should be preferred for assessment of soft-tissue injuries and oncologic imaging.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography