[Effect of Paroxysmal Atrial Fibrillation on Parameters Monitored by Transpulmonary Thermodilution Method]
Hua-dong Zhu1, Liang-liang Zheng1, Chen Li1
1Department of Emergency Medicine,PUMC Hospital,CAMS and PUMC,Beijing 100730,China;
Objective:
To investigate the effect of atrial fibrillation on the accuracy of parameters monitored by transpulmonary thermodilution method.
Methods:
Totally 12 patients from emergency intensive care unit with paroxysmal atrial fibrillation were enrolled. The hemodynamic parameters such as heart rate, mean arterial pressure, cardiac index, systemic vascular resistance index, intrathoracic blood volume index, and extravascular lung water index were monitored by transpulmonary thermodilution method before paroxysmal atrial fibrillation and during atrial fibrillation, the number of B-lines was detected by lung ultrasonography before and during paroxysmal atrial fibrillation. The changes of all the parameters were analyzed.
Results:
When the paroxysmal atrial fibrillation happened, the heart rate increased significantly [(123.3±20.0) beat/min vs. (98.9±12.3) beat/min, P=0.006]; the mean arterial pressure [(86.9±10.2) mmHg vs. (93.0±12.5) mmHg, P=0.058], cardiac index [(2.82±0.62) L/(min·m(2)) vs. (3.31±1.02) L/(min·m(2)), P=0.058] and systemic vascular resistance index [(2254±947) dyn·s·cm(-5)·m(2) vs. (2302±828) dyn·s·cm(-5)·m(2), P=0.351] had no obvious change; however, the intrathoracic blood volume index significantly increased [(1333±90) ml/m(2) vs. (937±111) ml/m(2), P<0.001]; extravascular lung water index also increased significantly [(16.1±1.1) ml/kg vs. (6.5±1.9) ml/kg, P<0.001]. No significant difference was found in the number of B-lines detected by lung ultrasonography before and during atrial fibrillation (10.0±4.2 vs. 9.4±4.4, P=0.180).
Conclusion:
Both intrathoracic blood volume and extravascular lung water monitored by transpulmonary thermodilution method were overvalued during paroxysmal atrial fibrillation, which may mislead the clinical judgment and decision-making.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
08:12Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
Related Concept Videos
Measurement of Blood Pressure
Dysrhythmias III: Characteristics of Dysrhythmias