Parametric electrical impedance tomography for measuring bone mineral density in the pelvis using a computational
Shani Kimel-Naor1, Shimon Abboud1, Marina Arad2
1Department of Biomedical Engineering, Faculty of Engineering, Tel-Aviv University, Tel-Aviv 69978, Israel.
Medical Engineering & Physics
|May 18, 2016
Summary
Parametric electrical impedance tomography (pEIT) shows promise for monitoring bone mineral density (BMD) changes associated with osteoporosis. This non-invasive technique could offer an accessible tool for frequent BMD assessments in clinical settings.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Osteoporosis Research
Background:
- Osteoporosis is characterized by bone microstructure deterioration, decreasing bone strength.
- Bone mineral density (BMD) is crucial for diagnosing osteoporosis, managing patients, and assessing fracture risk.
- Current BMD measurement methods like DEXA can be expensive and less accessible for frequent monitoring.
Purpose of the Study:
- To evaluate the efficacy of parametric electrical impedance tomography (pEIT) for monitoring BMD.
- To assess pEIT's potential as a complementary tool for frequent BMD assessments.
- To investigate the feasibility of using pEIT in clinical settings for osteoporosis management.
Main Methods:
- Developed a numerical solver to simulate surface potentials over a 3D pelvic model.
- Simulated varying cortical and cancellous BMD by altering bone conductivity and permittivity.
- Validated numerical simulations with preliminary measurements from a healthy male subject using a bioimpedance device.
Main Results:
- Simulated BMD changes from normal (100%) to osteoporotic (60%) resulted in up to 35% change in the real and 16% in the imaginary modules of calculated potentials.
- Negligible BMD relative error was achieved with a signal-to-noise ratio (SNR) greater than 60 dB.
- Positioning accuracy is critical for reliable long-term monitoring.
Conclusions:
- pEIT demonstrates potential for detecting BMD variations indicative of osteoporosis.
- The method shows promise as an inexpensive and user-friendly tool for frequent BMD monitoring.
- pEIT could serve as a complementary method to DEXA scans, particularly in small clinics during treatment.