Role of Optical Coherence Tomography in Assessing Anterior Chamber Angles
Reema Thomas Kochupurakal1, Krishnagopal Srikanth2, Kirti Nath Jha3
1Resident, Department of Ophthalmology, Mahatma Gandhi Medical College and Research Institute, Sri Balaji Vidyapeeth University , Pondicherry, India .
Spectral Domain Optical Coherence Tomography (SD-OCT) shows high sensitivity but low specificity for detecting narrow anterior chamber angles compared to gonioscopy. SD-OCT may be unreliable for screening narrow angles.
Area of Science:
- Ophthalmology
- Medical Imaging
- Anatomy
Background:
- Gonioscopy is the standard for assessing anterior chamber angles but suffers from interobserver variability.
- There is a need for a reliable, objective method for anterior chamber angle assessment.
Purpose of the Study:
- To compare anterior chamber angle measurements between gonioscopy and Spectral Domain Optical Coherence Tomography (SD-OCT).
- To evaluate SD-OCT's utility in individuals with shallow anterior chambers.
Main Methods:
- A comparative observational study included 101 eyes of 54 patients with shallow anterior chambers.
- Anterior chamber angles were graded using gonioscopy (Shaffer system) and SD-OCT (measuring Trabecular Iris Angle - TIA and Angle Opening Distance - AOD).
- Statistical analysis included Chi-square test, sensitivity, specificity, and predictive values.
Main Results:
- SD-OCT identified narrow angles (TIA ≤ 22°, AOD ≤ 290 μm) in a high percentage of eyes across quadrants.
- Sensitivity for detecting narrow angles was high (90.7% for TIA, 82.2% for AOD).
- Specificity was low (11.7% for TIA, 23.4% for AOD) when compared to gonioscopy.
Conclusions:
- SD-OCT identified narrow angles more frequently than gonioscopy.
- The high sensitivity and low specificity of SD-OCT suggest it may be unreliable for screening narrow anterior chamber angles.
- Further research may be needed to refine SD-OCT's role in angle assessment.