Related Experiment Video
Updated: Mar 20, 2026

Autologous Endothelial Progenitor Cell-Seeding Technology and Biocompatibility Testing For Cardiovascular Devices in Large Animal Model
Published on: September 9, 2011
Regulatory approval of new medical devices: cross sectional study
Hani J Marcus1, Christopher J Payne2, Archie Hughes-Hallett2
1The Hamlyn Centre, Institute of Global Health Innovation, Imperial College, London W2 1NY, UK Department of Neurosurgery, Imperial College Healthcare NHS Trust, London, UK hani.marcus10@imperial.ac.uk.
Objective:
To investigate the regulatory approval of new medical devices.
Design:
Cross sectional study of new medical devices reported in the biomedical literature.
Data Sources:
PubMed was searched between 1 January 2000 and 31 December 2004 to identify clinical studies of new medical devices. The search was carried out during this period to allow time for regulatory approval.
Eligibility Criteria For Study Selection:
Articles were included if they reported a clinical study of a new medical device and there was no evidence of a previous clinical study in the literature. We defined a medical device according to the US Food and Drug Administration as an "instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article."
Main Outcome Measures:
Type of device, target specialty, and involvement of academia or of industry for each clinical study. The FDA medical databases were then searched for clearance or approval relevant to the device.
Results:
5574 titles and abstracts were screened, 493 full text articles assessed for eligibility, and 218 clinical studies of new medical devices included. In all, 99/218 (45%) of the devices described in clinical studies ultimately received regulatory clearance or approval. These included 510(k) clearance for devices determined to be "substantially equivalent" to another legally marketed device (78/99; 79%), premarket approval for high risk devices (17/99; 17%), and others (4/99; 4%). Of these, 43 devices (43/99; 43%) were actually cleared or approved before a clinical study was published.
Conclusions:
We identified a multitude of new medical devices in clinical studies, almost half of which received regulatory clearance or approval. The 510(k) pathway was most commonly used, and clearance often preceded the first published clinical study.
More Related Videos
Related Concept Videos
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Impact of Pharmacokinetic–Pharmacodynamic Models: Regulatory Decisions
FDA Approved Drugs: Changes to Approved Drugs
Bioequivalence studies: Biowaivers
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...