Related Experiment Video
Updated: Mar 20, 2026

A Dual-Functional Electroactive Filter Towards Simultaneously SbIII Oxidation and Sequestration
Published on: December 5, 2019
Determination of Sb(III) using an ex-situ bismuth screen-printed carbon electrode by adsorptive stripping voltammetry
Carlos Rojas-Romo1, Núria Serrano2, Cristina Ariño2
1Departament de Química Analítica, Facultat de Química, Universitat de Barcelona, Martí i Franquès 1-11, E-08028 Barcelona, Spain; Facultad de Química, Pontificia Universidad Católica de Chile, Vicuña Mackenna 4860, Santiago 7820436, Chile.
Abstract:
The determination of Sb(III) on an ex-situ bismuth screen-printed carbon electrode (ex-situ BiSPCE) by means of adsorptive stripping voltammetry (AdSV) using quercetin-5'-sulfonic acid as chelating agent was optimized. The effect of different experimental parameters such pH, ligand concentration (CQSA), accumulation potential (Eacc) and accumulation time (tacc) were studied to obtain a wide linear range, the highest sensitivity and the lowest detection limit. Ex-situ BiSPCE was analytically compared with a sputtered bismuth screen-printed electrode (BispSPE) under optimal conditions. The obtained analytical parameters suggest that ex-situ BiSPCE behaves much better than BispSPE and the first was selected for this study. Optimal parameters were pH=4.6; CQSA=10.0 to 20.0×10(-6)molL(-1); Eacc=-0.5V and tacc=60s. Peak area is proportional to Sb(III) concentration up to 100.0μgL(-1) (tacc 60s) and 45.0μgL(-1) (tacc 120s) range, with detection limits of 1.2μgL(-)(1) (tacc 60s) and 0.8μgL(-1) (tacc 120s). The relative standard deviation for a Sb(III) solution (20.0μgL(-1)) was 3.9% for ten successive assays. Thus, the effect of various interfering metal ions was studied and the methodology was validated using a spiked groundwater reference material with very satisfactory results.
More Related Videos
13:09Assessment of Boron Doped Diamond Electrode Quality and Application to In Situ Modification of Local pH by Water Electrolysis
Published on: January 6, 2016
08:03Multi-analyte Biochip MAB Based on All-solid-state Ion-selective Electrodes ASSISE for Physiological Research
Published on: April 18, 2013
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Electrodeposition
Electrodeposition can...
Ion-Exchange Chromatography