Development and Validation of a Rapid RP-HPLC-DAD Analysis Method for the Simultaneous Quantitation of Paclitaxel and
Ebrahim Saadat1, Fatemeh Ravar1, Pouya Dehghankelishadi2
1Department of Pharmaceutics, Faculty of Pharmacy, Tehran University of Medical Sciences, Tehran P.O. Box 14155-645, Iran.
Abstract:
A robust and rapid analysis method was developed and validated for the simultaneous assay of paclitaxel (PTX) and lapatinib (LPT) in a polymeric micelle formulation as a novel drug delivery system using high-performance liquid chromatography (HPLC). The assay was performed using the C18 MZ-Analytical Column (5 μm, 150 × 4.6 mm, OSD-3) which was protected with the C18 pre-column (5 μm, 4.0 × 4.6 mm, OSD-3). The mobile phase was composed of acetonitrile and water (70/30; V/V) with a flow rate of 0.5 mL/min and detection wavelength of 227 nm. Accuracy was reported as the relative error and was found to be less than 6.8%. The interday assay was evaluated to be 3.22% and 5.76% RSD for PTX and LPT, respectively. The intraday precision was found to be at its maximum value of 5.83% RSD. The limit of detection for both PTX and LPT was found to be 1 µg/mL by means of the newly developed method. The limit of quantitation for PTX and LPT was found to be 5 µg/mL. The calibration curves for both drugs were linear in the concentration range of 5 to 80 μg/mL. In vitro release for both drugs from the polymeric micelle was evaluated using the newly developed analysis method.