Related Experiment Video
Updated: Mar 20, 2026

Microfluidics in Assessing Platelet Function
Published on: November 8, 2024
A Novel Device for the Evaluation of Hemostatic Function in Critical Care Settings
Elisa A Ferrante1, Kiev R Blasier, Thomas B Givens
1From *HemoSonics LLC, Charlottesville, Virginia; and †University of Virginia, Charlottesville, Virginia.
Abstract:
Major surgical procedures often result in significant intra- and postoperative bleeding. The ability to identify the cause of the bleeding has the potential to reduce the transfusion of blood products and improve patient care. We present a novel device, the Quantra Hemostasis Analyzer, which has been designed for automated, rapid, near-patient monitoring of hemostasis. The Quantra is based on Sonic Estimation of Elasticity via Resonance Sonorheometry, a proprietary technology that uses ultrasound to measure clot time and clot stiffness from changes in viscoelastic properties of whole blood during coagulation. We present results of internal validation and analytical performance testing of the technology and demonstrate the ability to characterize the key functional components of hemostasis.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017