Alopecia associated with unexpected leakage from electron cone
B C Wen1, E C Pennington, D H Hussey
1Department of Radiology, University of Iowa College of Medicine, Iowa City 52242.
Abstract:
Excessive irradiation due to unexpected leakage was found on a patient receiving electron beam therapy. The cause of this leakage was analyzed and the amount of leakage was measured for different electron beam energies. The highest leakage occurred with a 6 x 6 cm cone using a 12 MeV electron beam. The leakage dose measured along the side of the cone could be as great as 40%. Until the cones are modified or redesigned, it is advised that all patient setups be carefully reviewed to assure that no significant patient areas are in the side scatter region.