Errors in 14 pulse oximeters during profound hypoxia.

J W Severinghaus1, K H Naifeh, S O Koh

  • 1Department of Anesthesia, University of California, San Francisco 94143-0542.

Journal of Clinical Monitoring
|April 1, 1989
PubMed
Summary

Pulse oximeter accuracy varies significantly between manufacturers, especially at low oxygen saturation levels. Many devices underestimate oxygen levels and show reduced precision, impacting patient monitoring.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Respiratory responses mediated through superficial chemosensitive areas on the medulla

Journal of applied physiology (Bethesda, Md. : 1985)·2019
Same author

The predictors for continuous renal replacement therapy in liver transplant recipients.

Transplantation proceedings·2014
Same author

Evaluation of acid-base status in brain dead donors and the impact of metabolic acidosis on organ retrieval.

Minerva anestesiologica·2013
Same author

Randomized controlled comparison of combined general and epidural anesthesia versus general anesthesia on diaphragmatic function after laparoscopic prostatectomy.

Minerva anestesiologica·2013
Same author

Low-dose remifentanil infusion during ventilator weaning and tracheal extubation in postoperative intensive care unit patients sedated with propofol-remifentanil: a randomised clinical trial.

Anaesthesia and intensive care·2012
Same author

Per cent pH regulation as an index of CO₂ buffering in vivo and in vitro.

Scandinavian journal of clinical and laboratory investigation·2010

Area of Science:

  • Biomedical Engineering
  • Medical Device Technology
  • Clinical Monitoring

Background:

  • Pulse oximetry is a critical, non-invasive method for monitoring oxygen saturation.
  • Variability in pulse oximeter performance can affect clinical decision-making.
  • Previous studies have indicated potential inaccuracies in pulse oximeter readings.

Purpose of the Study:

  • To evaluate and compare the accuracy of pulse oximeters from multiple manufacturers.
  • To identify performance differences, particularly under hypoxic conditions.
  • To quantify the bias and precision of various devices.

Main Methods:

  • Tested 14 manufacturers' pulse oximeters on 125 subject sets during induced hypoxic plateaus.
  • Collected data at varying oxyhemoglobin (O2Hb) levels, including profound hypoxia (40-70%) and specific regression points (approx. 50-86%).
  • Defined an "ambiguity" index (bias and precision) to quantify error, comparing SpO2 to arterial O2Hb analyzed by a reference oximeter.

Main Results:

  • Significant differences in bias and precision were observed among the 14 manufacturers.
  • Common issues included underestimation of oxygen saturation and decreased precision at lower saturation levels.
  • Ambiguity values varied widely, with some devices showing substantial errors (e.g., Ohmeda, Radiometer, Nellcor).

Conclusions:

  • Substantial performance variations exist between pulse oximeter brands, especially during hypoxia.
  • Underestimation of saturation and loss of precision are key problems at low O2Hb levels.
  • Clinical use requires careful consideration of device-specific accuracy limitations.