[Monitoring tidal volumes when using the Ventrain® emergency ventilator].
A R Schmidt1,2, K Ruetzler3,4, T Haas5,6
1Anästhesie-Abteilung, Universitäts-Kinderspital Zürich, Steinwiesstrasse 75, 8032, Zürich, Schweiz. alexander.schmidt@kispi.uzh.ch.
The Florian respiratory function monitor accurately measures tidal volumes during ventilation with the Ventrain® device. This new monitoring capability can help prevent patient trauma by ensuring adequate ventilation and preventing excessive air trapping.
Area of Science:
- Medical Devices
- Respiratory Care
- Clinical Engineering
Background:
- The Ventrain® device provides emergency ventilation but lacks tidal volume monitoring, risking patient trauma.
- Accurate monitoring of inspiratory and expiratory tidal volumes is crucial for safe ventilation, especially in cases of airway obstruction.
Purpose of the Study:
- To evaluate the efficacy and precision of the Florian respiratory function monitor (RFM) for measuring tidal volumes delivered by the Ventrain® emergency ventilation device.
- To assess the RFM's ability to monitor both inspiratory and expiratory tidal volumes in an in-vitro setting.
Main Methods:
- An in-vitro study using the ASL 5000 test lung and a 2mm transtracheal catheter.
- The Florian RFM with its neonatal flow sensor was used to measure tidal volumes at various driving flows.
- Calibration and bias correction were performed, followed by comparison with ASL 5000 measurements using Bland Altman analysis.
Main Results:
- The Florian RFM, after software adaptation, showed a mean deviation of 2.59% for inspiratory and 1.66% for expiratory tidal volumes compared to the ASL 5000.
- Bias and precision were within acceptable limits, with bias of 2.05ml (23.20) for inspiration and 4.62ml (10.40) for expiration.
Conclusions:
- The Florian RFM, utilizing hot-wire anemometer technology, can effectively monitor tidal volumes delivered by the Ventrain®.
- Software adaptation allows the RFM to measure precise inspiratory and expiratory tidal volumes, aiding in adequate patient ventilation and reducing trauma risk.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Respiratory Volumes and Capacities